Running an experiment¶
The golden-path sequence. If the run does not follow this pattern, stop and re-read the Pre-run checklist.
1. Pre-run checklist complete¶
All items in Pre-run checklist are checked off. Stop-path verification done if this is the first run of the day.
2. Start the laser (if flow-vis)¶
- Confirm 12 V on the power supply.
- Enable the supply's output.
- Confirm the laser sheet appears in the test section and terminates on the dump surface.
- Confirm door placard shows laser in use.
3. Start the tunnel at a low RPM first¶
Bring the tunnel up gradually; do not jump to high RPM from idle.
- In ADCL WinSoft, Quick Tunnel panel: set the slider to 100 RPM (or whatever your run plan's lowest point is).
- Press Apply Setpoint. The drive accepts REF1.
- Look at the test section one more time. Is it clear of people and tools?
- Press Start Tunnel. The drive transitions through the start sequence. The motor spins up.
- Confirm the Speed RPM readout settles at the commanded value (100 RPM).
4. Start the DAQ (if recording)¶
- Confirm the footer shows DAQ: cDAQ1.
- In the right column, press Start DAQ.
- The Signals tab shows live readings.
- The CSV is being written to
E:\Wind_tunnel\ADCLWinSoft\Data\YYYY-MM-DD\….
5. Start the smoke generator (if flow-vis)¶
The tunnel must already be running. Smoke without airflow pools in the room.
- Confirm the wired remote is at your station.
- Press the smoke generator's ON button. A puff of smoke enters the inlet.
- Adjust the heater / fan settings if needed for the visualization quality you want.
6. Run the experiment¶
- Adjust the Quick Tunnel slider to each target RPM in your run plan.
- Press Apply Setpoint to write the new RPM. Do not press Start Tunnel again — that stops and restarts the drive.
- Wait for the speed reading to stabilize at each setpoint before taking data.
- Watch the footer for:
- Connected (Modbus, green)
- DAQ: cDAQ1
- No red Fault banner
- If a fault appears, stop the run and consult a contributor.
7. Stop the smoke generator (if flow-vis)¶
Before stopping the tunnel:
- Press the smoke generator OFF.
- Wait until the visible plume in the test section clears (10–20 s of continued airflow).
8. Ramp-stop the motor¶
- Press Stop Tunnel in ADCL WinSoft. The motor decelerates per the drive's ramp time.
- Confirm the Speed RPM readout reaches 0 and the drive keypad shows 0 Hz.
9. Stop the DAQ recording¶
- In the right column, press Stop DAQ.
- The CSV is finalized. Note its filename in your run log.
10. Stop the laser¶
- Disable the supply's output.
- Power the supply down.
11. Continue or shutdown¶
- To run another experiment immediately, return to step 1 and repeat the pre-run checklist (an abbreviated version is fine if nothing in the lab has changed).
- To end the session, follow the Shutdown checklist.
Do not, ever¶
- Press Start Tunnel while the motor is already running. This stops it.
- Reach into the test section while the motor is energized, even at zero commanded RPM.
- Reach into the exhaust path while the motor is energized.
- Leave the computer station while the motor is running.
- Use AeroWare to spin the motor (its VFD path is broken; see the contributor manual).
- Increase the laser power supply above 12 V.